Ovation Ult Specifications

Body Balance OvationULT: Size, Weight and Room Fit

  • Model: OvationULT Zero Gravity Red Light Therapy Bed
  • Bed size (L x W x H): 80 x 31 x 40 in
  • Bed weight: 175 lb, two-person carry
  • Client weight limit: 450 lb
  • Door needed: Standard 32 in opening, no disassembly
  • Smallest room: 10 x 10 ft
  • Delivery state: Ships fully assembled
  • Location: Indoor use only

OvationULT light output

  • Total diodes: 28,443
  • Red diodes: 22,755 at 635 nm
  • Near-infrared diodes: 5,688 at 850 nm
  • Irradiance: 65 mW/cm², plus or minus 10, at the polycarbonate surface
  • Lens spacing: 9/16 in from center
  • Light to body: Canopy rests on the skin, 0 to 2 in
  • Stated diode life: 40,000 hours average, about 8 to 10 years of use

OvationULT sessions and controls

  • Session timer: 10, 15 or 20 minutes
  • Entry delay: 15 seconds after a timer is selected
  • Massage: Vibration frequency massage, switched on with its own button
  • Stop mid-session: Press Power to turn the lights off
  • Access: Client enters and exits alone

OvationULT power and materials

  • Outlet: Standard 3-prong
  • Power input: 115/230 VAC
  • Power output: 24 VDC, 18.8 A
  • Stated power: Over 332 W
  • Surface: Polycarbonate, ISO 10993-1 and USP Class VI
  • FDA status: FDA registered Class II device, product code ILY, registration 3010627475
  • Made in: Las Vegas, Nevada, USA

How a session runs on the OvationULT

  1. Press Power. Press Massage if the client wants it.
  2. Select 10, 15 or 20 minutes. The client has 15 seconds to get in.
  3. The client lies back and pulls the canopy down until it touches the skin, arms outside the side panels.
  4. Lights turn off when the timer ends. The client pushes the canopy up and steps out.
  5. Wipe the polycarbonate with disinfectant on a microfiber cloth.

Eye protection is optional for comfort. Clients pull the canopy by its center, never by the side panels.

Client lying in the Body Balance OvationULT with the canopy lowered

Cleaning products for the OvationULT

Use disinfectant with hydrogen peroxide or isopropyl alcohol. Avoid bleach, ammoniated chloride, and products based on aldehydes, phenols, hydroxides or amines.

Clients who need a doctor's approval first

Body Balance lists epilepsy, pregnancy or breastfeeding, liver or kidney disorders, pacemakers, uncontrolled high blood pressure, severe thyroid problems, cancer, a weakened immune system, and heart disease or arrhythmia. Stop any session that causes pain or discomfort.

OvationULT manuals and documents